Technical data sheet

Getinge GE 2612 ARB-2 steam sterilizer

Ref. PP-1037 · 2000

Technical data

ManufacturerGetinge
Model / typeGetinge GE 2612 ARB-2
Year of manufacture2000
ConditionOverhauled
Serial number5100810-060-01
Capacity / output564 l chamber / 108 l jacket
Power supply400 V, 3 Ph, 50 Hz · 2,5 kW · 6,1 A · Steuerspannung 24 V AC
GMP / documentation statusComplete documentation available

Documentation

Operating manual, Electrical drawings, Spare parts list, Pressure vessel certificate

Description

Getinge GE 2612 ARB-2 steam sterilizer, built in 2000 by Getinge AB, serial no. 5100810-060-01. Chamber volume 564 litres, jacket 108 litres; design pressure 3 / −1 bar in the chamber and 3.5 / −1 bar in the jacket, test pressure 3.9 and 5.9 bar respectively, design temperature 148 °C for both. The process control was renewed in 2020. Electrical supply 3~50 Hz, 400 V, rated 2.5 kW at 6.1 A, control voltage 24 V AC; the original electrical diagram (drawing 565 53 04) is available. There is surface corrosion on the outer frame above the nameplate, visible in the photographs. Available for inspection by appointment, sold ex works in the condition inspected.

Price on request